3-8

KLN Series 051614

3.3.1.1

FRONT PANEL LOCKOUT

To prevent unauthorized setting of the power supply, KEYLOCK can be enabled which disables
the front panel keys and the encoder.

Press SHIFT key to enter SHIFT mode (

LED lit), then tap the encoder to enable KEYLOCK.

The yellow

indicator goes on and all the front panel keys (except SHIFT) and the encoder

are disabled.

To exit KEYLOCK tap the encoder while in SHIFT mode (

LED lit).

3.3.2

SET VOLTAGE OR OVERVOLTAGE PROTECTION (OVP)

The maximum setting for voltage is 105% of rated voltage and cannot be adjusted to exceed the
OVP setting. The maximum setting for OVP is 110% of rated voltage and cannot be set to less
than the programmed value of V. See PAR. B.32 and B.36 for SCPI commands.

1. Verify Output Mode set to ENTER (see PAR. 3.3.8)

2. To set voltage, press V/OVP (V set) key. To set OVP, first press SHIFT key (blue

LED

goes on), then press V/OVP key.

3. The right-most digit blinks. Rotate the encoder to change the value. Tap the encoder to go to

the next digit. When adjustment is complete, press ENTER to confirm.

3.3.3

SET CURRENT OR OVERCURRENT PROTECTION (OCP)

The maximum setting for current is 105% of rated current and cannot be adjusted to exceed the
OCP setting. The maximum setting for OCP is 110% of rated current and cannot be set to less
than the programmed value of A. See PAR. B.14 and B.16 for SCPI commands.

1. Verify Output Mode set to ENTER (see PAR. 3.3.8)

2. To set current, press A/OCP (A set) key. To set OCP, first press SHIFT key (blue

LED

goes on), then press A/OCP key.

3. The right-most digit blinks. Rotate the encoder to change the value. Tap the encoder to go to

the next digit. When adjustment is complete, press ENTER to confirm.

CC.M

LOC or EXT

Current set, either LOC (Local) using the front panel or EXT (External)
using the analog signal from the Programming Control port.

PAR. 3.3.40

CAL

0000

Calibration password.

PAR. 3.3.41

SNnn

nnnn

Serial Number, where nnnnnn = 6 numbers, two numbers in Voltage
Display, four numbers in Current display.

PAR. 3.3.42

FW

x.xx

Firmware version where xx.x = numbers (cannot be modified)

PAR. 3.3.43

HW

y.yy

Hardware version where y.yy = numbers.

PAR. 3.3.44

V.MA1

(2)

70.46

View part 1 of MAC Address

PAR. 3.3.45

V.MA2

(2)

42.xx

View part 2 of MAC Address

PAR. 3.3.46

V.MA3

(2)

xx.xx

View part 3 of MAC Address

PAR. 3.3.47

RST

OFF or ON

Reset Function.

PAR. 3.3.48

RSTI

(2)

OFF or ON

IP Address Reset Function.

PAR. 3.3.49

(1) Only if optional GPIB interface installed. (2) Only if optional LAN interface installed.

TABLE 3-3. MENU FUNCTIONS

Voltage Display

Current Display

Description

Reference