KEPCO BOP 100-2DC Modification Sheet User Manual
Kepco, Modification of standard
The Kepco Models BOP 100-2MC and BOP 100-2DC have been modified from standard Kepco
Models BOP 100-2M and BOP 100-2D, respectively, to be stable handling capacitive loads up
to 10 mF.
INSTRUCTION MANUAL CORRECTIONS:
This modification makes the BOP more suitable for a wide variety of applications such as
solar cell/ solar panel testing, piezoelectric device driving/ testing, capacitor testing, capacitive
transducer driving/testing and powering industrial or lab-type applications for capacitive or
capacitive-resistive loads.
Static specifications representing the unit’s accuracy in Voltage mode are identical with the
standard BOP models, while the ripple and noise specifications are better (approximately 50%
lower) for MC and DC models than standard BOP models.
Specifications listed in Table 1 are for BOP 100-2MC and BOP 100-2DC in Voltage Mode.
In voltage mode the bandwidth of the BOP 100-2MC and DC models is less than the stan-
dard BOP, while the response time is increased. Nonuniformities of the frequency response can
be nearly eliminated by reducing the bandwidth in Voltage mode. Bandwidth can be reduced in
a predictable manner by increasing the internal compensation capacitance by placing an exter-
nal capacitor in parallel across C21 of Assembly A1 (see Table 2 and Figure 1).
TABLE 1. BOP 100-2MC AND BOP 100-2DC SPECIFICATIONS
Bandwidth (DC to f-3dB)
Rise/Fall Time
(2)
Recovery at Step Load
(3)
Resistive Load, Nominal
Capacitive Load, 10
µF
(1)
5.0 kHz
5.6 kHz
74
µ
S
115
µ
S
NOTES:
(1) Nonuniformities of the frequency response for the standard 10
µF load create a larger 3-dB bandwidth than
for the resistive load.
(2) 10% to 90%, with nominal resistive load.
(3) Load between infinity and nominal resistive load values.
TABLE 2. BANDWIDTH CORRECTION
EXTERNAL CAPACITOR
(ACROSS A1C21, SEE FIGURE 1)
1 nF
2.2 nF
4.7 nF
15 nF
33 nF
47 nF
100 nF
CORRECTED BANDWIDTH
4.9 kHz
4.7 kHz
4.1 kHz
2.7 kHz
1.4 kHz
1.05 kHz
0.65 kHz
NOTE: Listed bandwidth correction values are for Voltage mode, nominal resistive load.
KEPCO
®
THE POWER SUPPLIER™
MODIFICATION OF STANDARD
KEPCO MODEL BOP 100-2MC
KEPCO MODEL BOP 100-2DC
BOP 100-2MC/DC-r1 031512
KEPCO, INC. z 131-38 SANFORD AVENUE z FLUSHING, NY. 11355 U.S.A. z TEL (718) 461-7000 z FAX (718) 767-1102