beautypg.com

6 pr-23-ac i-line fitting, 7 3a sanitary standard compliance, 9 sensor specifications 83 – K-Patents PR-23 User Manual

Page 91: Figure 9.9 i-line fitting for pr-23-ac

background image

9 Sensor specifications

83

9.3.6 PR-23-AC I-Line fitting

K-Patents Sanitary Refractometer PR-23-AC-ZC can be mounted using 3-A Sanitary
approved 2.5 inch Cherry Burrell I-Line fittings that are made of interlocking flat face
ferrules, a flat gasket and a clamp. This interlocking, metal-to-metal design eliminates
over compression by the clamp not allowing the gasket to be extruded into the product
contact side.

Sensor wetted parts material is AISI 316L or Hastelloy C, gaskets EPDM.

B

B

SECTION B-B

1

2

3

4

Flow

#

Desc

Material

Pcs

1

PR-23-AC-Z

***

1

2

I-line gasket

EPDM

1

3

I-line clamp

Aisi 304

1

4 Flowcell 3" Pipe bend,

2833

Aisi 316L

1

Appr.

Drawn

Drawing number

Drawing description

Mass

General tolerances

MTG

kg

07.08.2013 AHä

07.08.2013 TL

ISO 2768 FINE

Revision

2825

PRE.01

Scale

1:2

I-line flowcell 3" pipe bend

Material

Figure 9.9

I-Line fitting for PR-23-AC

9.3.7 3A Sanitary Standard compliance

The user should ensure that the refractometer is not a source of contamination to
product due to damaged or worn product contact surfaces. Misuse (e.g. too long prism
wash time or too high wash pressure) or mishandling may result in metal scratches
or roughened surfaces. Such surfaces may not stay clean in processing.

K-Patents offers a 3A Sanitary Standard Accepted repair and maintenance package in
which all wetted parts, prism, gaskets and dryer are replaced. Note that this repair
service can be completed only by a 3A authorized service center (K-Patents factory or
regional headquarters).