beautypg.com

K-Patents PR-23 User Manual

Page 58

background image

50

PR-23 instruction manual

For a complete report, 10–15 valid data points (see below) are needed. A data point
is of use for calibration only when the diagnostic message is

NORMAL OPERATION

. If prism

wash is employed, do not take samples during the wash. Each data point consists of:

LAB%

Sample concentration determined by the user

From DTR display:

(see Figure 6.13)

CALC

Calculated concentration value

T

Process temperature measurement in Centigrade

nD

Actual refractive index n

D

CONC

Measurement in concentration units, the large size number

In addition to the calibration data, write down the Indicating transmitter serial num-
ber, the sensor serial number and the sensor position, i.e. whether it is installed as
Sensor A or as Sensor B.

Accurate calibration is only achieved if the sample is taken correctly. Pay special at-
tention to following details:

− The sampling valve and the refractometer should be installed close to each other

in the process.

Warning! Wear protective clothing appropriate to your process when operating the
sampling valve and handling the sample.

− Run the sample before starting to collect data points to avoid sampling old process

liquid that has remained in the sampling valve.

− Read the values

CALC

,

T

(emp),

nD

and

CONC

in the DTR’s display at exactly the same

time with sampling.
The easiest way of doing this is to use the

FIELD SAMPLE

soft key available through

the Sensor status display (DTR program version 2.0 or newer). The value of each
sample is the average of 10 consequent measurements to increase accuracy and
reduce possible process noise.

Press the

FIELD

SAMPLE

soft key...

...wait while

DTR measures...

(take sample for lab)

...and press

REPEAT

for next

data point or BACK to get to

the Sensor status display.

Figure 6.13

Using

FIELD SAMPLE

soft key

− Use a tight container for the sample to avoid evaporation.