beautypg.com

6 wash nozzle insertion and removal, 1 wash nozzle insertion – K-Patents PR-23 User Manual

Page 170

background image

162

PR-23 instruction manual

11.6 Wash nozzle insertion and removal

11.6.1 Wash nozzle insertion

Check the nozzle and the valve before installing the wash nozzle. Use thread seal
tape for all thread connections.
Warning! Always shut the main steam valve before performing any work on the
wash nozzle.

1

1

3

2

!

1. Insert the nozzle into the isolation

valve (1).

2. Attach the nozzle to the nozzle guide

with one special M5x10 screw (2) using a

5 mm allen key.

3. Remove the safety pin (3).

2

1

5

4

3

2

CLOSE

OPEN

1. Close the 1/4’’ check port valve under

the nozzle isolation valve (1).

2. Open the isolation valve (2) by turning

the handle counterclockwise.

3. Push the nozzle to the process (3).

4. Attach the nozzle to the nozzle guide

with one M5x10 screw (4) using a 4 mm

allen key.

5. Lock the isolation valve handle with the

safety pin (5).

3

OPEN

1

2

3

1. Connect the steam line and sensor

flush flexible line to the nozzle T-piece

(1,2).

2. Open the steam supply line valve (3).