9 Sensor specifications

117

9.8.5 PR-23-MS specifications

General specifications
Refractive Index range, standard:

Full range n

D

1.3200–1.5300 (corresponds to

hot water – 100% b.w.) with Sapphire prism
H73

Refractive Index range, option

With sapphire prism H74 n

D

1.2600–1.4700

Accuracy:

Refractive index n

D

± 0.0002 (corresponds typ-

ically to ± 0.1 % by weight)
Repeatability and stability correspond to accu-
racy

Speed of response:

1 s undamped, damping time selectable up to
5 min

Calibration:

With Cargille certified refractive index liquids
over full range of n

D

1.3200–1.5300

CORE-Optics:

No mechanical adjustments (US Patent No.
US6067151)

Digital measurement:

3648 pixel CCD element

Light source:

Light emitting diode (LED) 589 nm wave-
length, sodium light

Temperature sensor:

Built-in Pt-1000

Temperature compensation:

Automatic, digital compensation

Instrument verification:

With certified refractive index liquids and
K-Patents documented procedure

Ambient temperature:

Sensor: max. 45 °C (113 °F),
min. -20 °C (-4 °F)
Indicating transmitter: max. 50 °C (122 °F),
min. 0 °C (32 °F)

SENSOR PR-23-MS:

Teflon body sensor model for aggressive medium

Sensor protection class:

IP67, Nema 4X

Sensor weight:

5.0 kg (12.1 lbs)

FLOW CELL FOR PR-23-MS
Process connection:

Thread G 1/2" or 1/2" NPT female

Process pressure:

max. 10 bar (145 psi)

Process temperature:

max. 130 °C (266 °F)

Process wetted parts, standard:

Teflon

®

(PTFE) or Kynar

®

(PVDF), prism gas-

kets MTF (Modified Teflon

®

), prism Spinel, O-

ring Kalrez, adaptor sapphire

Process wetted parts, option:

prism sapphire