beautypg.com

K-Patents PR-23 User Manual

Page 169

background image

11 Safe-Drive™

161

10

Take a firm grip on the hand wheel and the handle

and pull out the Safe-Drive™ Retractor with the

sensor inside.

A firm hold of the tool is essential as the

combination of the tool and the sensor is noticeably

heavier than Retractor alone.

Note:

To ensure the isolation valve after the

Safe-Drive tool with the sensor have been removed,

you can bolt a standard ANSI 1.5" 105 lbs blind

flange to it with 1/2" (M12) bolts and nuts.

A lock can be added to the isolation valve handle.

Warning! The sensor tip is hot and may be covered

with liquor. It is recommended to rinse the sensor

tip and the isolation valve with hot water.

11

1

2

11

Put the Safe-Drive™ Retractor with sensor onto a

table or similar surface so that the hand wheel has

space to turn.

1 Turn the wheel counterclockwise until the

trapezoidal thread is all the way inside the Outer

casing and loose from the wheel, i.e. the Outer

casing is no longer connected to the parts inside.

2 Pull the Outer casing off.

12

1

2

3

12

1 Open the latch on the Inner casing.

2 Keep sensor steady with one hand and turn the

Inner casing counterclockwise with the other hand

to unlock the Inner casing from sensor.

3 Pull off the sensor.