beautypg.com

K-Patents PR-23 User Manual

Page 187

background image

13 Sensor verification

179

Figure 13.2

Verification, pre-verificaton checks

When you are finished with the preparations, press

CONTINUE

(right-most soft key)

to start the verification process.
The verification itself is done by the refractometer system, you only have to follow
the instructions on screen and apply one RI liquid at a time on the sensor and press

VERIFY

(right-most soft key). See Figure 13.3.

Figure 13.3

Verification display

Important:

Clean and dry the prism and the sample holder very carefully be-

tween the RI liquids. Use a suitable solvent, e.g., ethanol.
To check that the standard liquid is properly wetting the prism, it is possible to
press soft key

OPT. IMAGE

. The optical image should show a sharp shadow edge, as e.g.

in Figure 13.4. For more information on the optical image, please see section 5.4.1.

Sharp optical image (with IDS)

Sharp optical image (without IDS)

Figure 13.4

Typical optical images