K-Patents PR-23 User Manual
Page 86
78
PR-23 instruction manual
9.3.3 PR-23-AC specifications
General specifications
Refractive Index range:
Full range n
D
1.3200–1.5300 (corresponds to
hot water – 100 Brix)
Accuracy:
Refractive index n
D
± 0.0002 (corresponds typ-
ically to ± 0.1 % by weight)
Repeatability and stability correspond to accu-
racy
Speed of response:
1 s undamped, damping time selectable up to
5 min
Calibration:
With Cargille certified refractive index liquids
over full range of n
D
1.3200–1.5300
CORE-Optics:
No mechanical adjustments (US Patent No.
US6067151)
Digital measurement:
3648 pixel CCD element
Light source:
Light emitting diode (LED) 589 nm wave-
length, sodium light
Temperature sensor:
Built-in Pt-1000
Temperature compensation:
Automatic, digital compensation
Instrument verification:
With certified refractive index liquids and
K-Patents documented procedure
Ambient temperature:
Sensor: max. 45 °C (113 °F),
min. -20 °C (-4 °F)
Indicating transmitter: max. 50 °C (122 °F),
min. 0 °C (32 °F)
SENSOR PR-23-AC:
Compact sensor model for small pipe lines
Process connection:
Sanitary 3A-clamp 2.5"; Varivent
®
in-line ac-
cess unit clamp DN65 or via elbow flowcell
(for line sizes of 2.5" and smaller)
Sanitary design:
3-A Sanitary Standard 46-03 approved and
EHEDG (European Hygienic Equipment De-
sign Group) tested
Process pressure:
Sanitary clamp max. 15 bar (200 psi) at 20 °C
(70 °F)/9 bar (125 psi) at 120 °C (250 °F)
Process temperature:
-20 °C–+130 °C (-4 °F–+266 °F)
Process wetted parts, standard:
AISI 316L stainless steel, prism spinel, prism
gaskets MTF (Modified Teflon)
Sensor protection class:
IP67, Nema 4X
Sensor weight:
2.0 kg (4.4 lbs)