beautypg.com

3 prism wash nozzles – K-Patents PR-23 User Manual

Page 33

background image

4 Prism wash systems

25

4.2.3 Prism wash nozzles

When selecting a wash nozzle for a compact refractometer, take into account both
the wash medium and the flowcell model: flowcells with larger pipe diameters need
longer wash nozzles. Figure 4.7 below shows a wash nozzle for a flowcell and gives
the measurements and part numbers for each nozzle type.

0.8

Mark

Rev.

Rev. by

Date

Appr.

Title/Name, designation
Scale

General tolerances

Part no.

Drawing no.

Join

File

Dimensions, model no.

Standard

Quantity

Drawing description

Drawing number

Drawn

Appr.

A

B

756/C

756.drw

SFS 4011 FINE

1 1

:

30.8.2004 HS
30.8.2004 HS

AFC nozzles

Material

AISI316L

C

MRP nrot muuttuneet

Flowcell -H10 or -H15

A

B

part nr

steam

50

4.0

PR-3365

water

75

2.5

PR-3369

pressurized water

75

1.5

PR-3368

Flowcell -H20 or -H25

A

B

part nr

steam

72

4.0

PR-3375

water

97

2.5

PR-3379

pressurized water

97

1.5

PR-3378

Figure 4.7

Wash nozzles for flowcell AFC-HSS-XXX-XX-NC

Figure 4.8 shows how the nozzle is mounted in a flow cell (-NC with stud for a wash
nozzle). Note: See Section 9.3.5 for more information on flowcells.

Figure 4.8

Process connection of a wash nozzle in a flow cell

For probe refractometers, select the wash nozzle according to wash medium and
refractometer model, see Table 4.1 below.