3 prism wash nozzles – K-Patents PR-23 User Manual
Page 33

4 Prism wash systems
25
4.2.3 Prism wash nozzles
When selecting a wash nozzle for a compact refractometer, take into account both
the wash medium and the flowcell model: flowcells with larger pipe diameters need
longer wash nozzles. Figure 4.7 below shows a wash nozzle for a flowcell and gives
the measurements and part numbers for each nozzle type.
0.8
Mark
Rev.
Rev. by
Date
Appr.
Title/Name, designation
Scale
General tolerances
Part no.
Drawing no.
Join
File
Dimensions, model no.
Standard
Quantity
Drawing description
Drawing number
Drawn
Appr.
A
B
756/C
756.drw
SFS 4011 FINE
1 1
:
30.8.2004 HS
30.8.2004 HS
AFC nozzles
Material
AISI316L
C
MRP nrot muuttuneet
Flowcell -H10 or -H15
A
B
part nr
steam
50
4.0
PR-3365
water
75
2.5
PR-3369
pressurized water
75
1.5
PR-3368
Flowcell -H20 or -H25
A
B
part nr
steam
72
4.0
PR-3375
water
97
2.5
PR-3379
pressurized water
97
1.5
PR-3378
Figure 4.7
Wash nozzles for flowcell AFC-HSS-XXX-XX-NC
Figure 4.8 shows how the nozzle is mounted in a flow cell (-NC with stud for a wash
nozzle). Note: See Section 9.3.5 for more information on flowcells.
Figure 4.8
Process connection of a wash nozzle in a flow cell
For probe refractometers, select the wash nozzle according to wash medium and
refractometer model, see Table 4.1 below.