To set a low temperature limit, choose

8 TEMP LIMIT VALUE °C

and enter the temperature

(in °C!) where the limit should be.

The empty pipe check prevents washing if message is

NO SAMPLE

, i.e. there’s no

process liquid in the pipe. To deactivate (or active) the empty pipe check, choose

9 EMPTY PIPE CHECK

and then the appropriate menu command.

To change the wash nD limit, select first

0 MORE ...

and then

1 WASH nD LIMIT

to set the

n

D

value to be used with the wash check functionality.

To set wash tolerance time, select first

0 MORE ...

and then

2 WASH TOLERANCE TIME

to set

the time during which a wash has to be accepted. If no wash is accepted during wash
tolerance time, the message becomes

PRISM WASH FAILURE

. The wash tolerance counter is

reset by a successful wash.