8 Troubleshooting

73

8.4 Diagnostic messages table

Important: The messages are listed in descending order of priority. Example: If
both

NO OPTICAL IMAGE

and

TEMP MEASUREMENT FAULT

are activated, only

NO OPTICAL IMAGE

will

be shown. The wash related messages have priority only during the wash cycle
(preconditioning–wash–recovery).

Certain malfunctions cause the mA measurement to return to the mA default output
value (see section 6.5.1), see the table below.

Returns to default mA

Message

Section

Conc

Temp

SHORT-CIRCUIT

8.1.6

X

X

NO SIGNAL

8.1.5

X

X

OUTSIDE LIGHT ERROR

8.2.1

NO OPTICAL IMAGE

8.2.2

X

TEMP MEASUREMENT FAULT

8.2.7

X

X

PRECONDITIONING

8.3

WASH

8.3

RECOVERING

8.3

HIGH SENSOR HUMIDITY

8.1.7

HIGH SENSOR TEMP

8.1.8

HIGH TRANSMITTER TEMP

8.1.9

LOW TRANSMITTER VOLT

8.1.10

EXTERNAL WASH STOP

8.3.5

LOW TEMP WASH STOP

8.3.6

NO SAMPLE/WASH STOP

8.3.7

EXTERNAL HOLD

8.3.1

NO SAMPLE

8.2.6

X

PRISM COATED

8.2.3

X

OUTSIDE LIGHT TO PRISM

8.2.4

LOW IMAGE QUALITY

8.2.5

PRISM WASH FAILURE

8.3.4

PRISM WASH WARNING

8.3.3

NO SENSOR

8.1.4

X

X

NORMAL OPERATION