8 Troubleshooting
73
8.4 Diagnostic messages table
Important: The messages are listed in descending order of priority. Example: If
both
NO OPTICAL IMAGE
and
TEMP MEASUREMENT FAULT
are activated, only
NO OPTICAL IMAGE
will
be shown. The wash related messages have priority only during the wash cycle
(preconditioning–wash–recovery).
Certain malfunctions cause the mA measurement to return to the mA default output
value (see section 6.5.1), see the table below.
Returns to default mA
Message
Section
Conc
Temp
SHORT-CIRCUIT
8.1.6
X
X
NO SIGNAL
8.1.5
X
X
OUTSIDE LIGHT ERROR
8.2.1
NO OPTICAL IMAGE
8.2.2
X
TEMP MEASUREMENT FAULT
8.2.7
X
X
PRECONDITIONING
8.3
WASH
8.3
RECOVERING
8.3
HIGH SENSOR HUMIDITY
8.1.7
HIGH SENSOR TEMP
8.1.8
HIGH TRANSMITTER TEMP
8.1.9
LOW TRANSMITTER VOLT
8.1.10
EXTERNAL WASH STOP
8.3.5
LOW TEMP WASH STOP
8.3.6
NO SAMPLE/WASH STOP
8.3.7
EXTERNAL HOLD
8.3.1
NO SAMPLE
8.2.6
X
PRISM COATED
8.2.3
X
OUTSIDE LIGHT TO PRISM
8.2.4
LOW IMAGE QUALITY
8.2.5
PRISM WASH FAILURE
8.3.4
PRISM WASH WARNING
8.3.3
NO SENSOR
8.1.4
X
X
NORMAL OPERATION