beautypg.com

4 mounting, 3 safe-drive™ retractor – K-Patents PR-23 User Manual

Page 155

background image

11 Safe-Drive™

147

11.3.3 Safe-Drive™ Retractor

The Safe-Drive™ Retractor consists of Inner casing and Outer casing. The Inner casing
is attached to the sensor flange with bayonet mounting. The Outer casing is attached
to the isolation valve body with bayonet mounting. When the hand-wheel is turned,
the Inner casing moves inside the Outer casing along the screw thread.

Locking latch

Inner casing

Bayonet mounting

Screw thread

Spill port

Bayonet mounting

Locking latch

Outer casing

Carrying handle

Hand-wheel

Figure 11.2

The Safe-Drive™ Retractor

11.4 Mounting

A standard Safe-Drive™ system delivery contains a Safe-Drive™ Sensor (PR-23-SD)
with a Indicating transmitter DTR, a Safe-Drive™ Isolation valve to be welded onto cus-
tomer’s pipe and a Safe-Drive™ Retractor for sensor insertion and removal. A welding
guide sticker is also provided for accurate cutting and welding.
By special order the Safe-Drive™ Isolation valve can also be welded to a suitable length
of pipe at the K-Patents factory to be part of the piping on site.

1000-1500 (3-5ft)

600-1000 [2-3ft]

700

Floor, work platform

Figure 11.3

Selecting the mounting location