Page 186

background image

178

PR-23 instruction manual

13.1 Refractive index n

D

verification

Before starting the verifications procedures, make sure that you have a K-Patents
PR-23 sample holder at hand. Also check the condition of your standard refractive
index liquids. You’ll also need a cleaning solution (ethanol) to clean the sensor
prism and the sample holder.
The sample holder keeps the sample on the prism surface and also keeps the ambi-
ent light out. The K-Patents universal sample holder PR-1012 (Figure 13.1) can be
used with any K-Patents PR-23 sensor (in PR-23-M only the top part of the sample
holder is needed).

Figure 13.1

The universal sample holder PR-1012

The verification of the PR-23 sensor calibration is made using a set of standard
refractive index liquids with the nominal values at 25 °C:

− 1.330

− 1.370

− 1.420

− 1.470

− 1.520

The accuracy of the certified standard refractive index liquids is ±0.0002 and they
can be traced back to national standards: N.I.S.T Standards # 1823 and # 1823 II.
The repeatability of PR-23 sensor, i.e. the deviation from the latest n

D

calibration,

is within ±0.0002.
As the specified accuracy of PR-23 is ±0.0002, then the representative level is the
sum of the three accuracy specifications, that is ±0.0004.
K-Patents provides a set of standard R.I. liquids, PR-2300, containing these five
liquids. The set can be ordered directly from K-Patents or through your nearest
K-Patents representative.

13.2 Verification procedure

To start the verification process, select

1 VERIFICATION

in the Main menu of your sen-

sor. The first verification display instructs you on the pre-verification procedure: