PR-23 instruction manual

6

Open and remove the four M12 nuts on the bolts

holding the sensor to the Isolation valve using a

19 mm or 3/4" wrench.

7

1 Turn the hand wheel clockwise until it stops to

remove the sensor from process.

2 At this stage about 140 mm (5.5") of the thread

should stick out from the middle of the wheel.

Warning! If you detect leaking, revert immediately

to the previous step. Do not continue removal until

the reason for leakage has been cleared and fixed.

8

1 Lift up the isolation valve handle locking plate.

2 Close the Isolation valve on by turning the

handle 90 degrees (a quarter turn) to the left.

Important:

The Isolation valve is properly closed

when the handle points away from the sensor and

the locking plate drops down over the handle.

3 Open the blow-out valve under the Isolation

valve for box cleaning to get rid of any process

liquid inside the isolation valve.

Warning! Some process liquid will leak out through

the small ball valve; beware of splashing!

9

1 Remove the safety pin.

2 Lift the Outer casing locking latch.

3 Turn the Outer casing 60 degrees

counterclockwise (to the left) so that the handle

comes up on top.