K-Patents PR-23 User Manual
Page 61

6 Configuration and calibration
53
The other
sensor
washing?
NO
YES
MANUAL
WASH
been
pressed?
NO
YES
Remote
wash
requested?
1
NO
YES
Wash
interval
reached?
YES
NO
Temp
limit
active?
YES
NO
Temp
above
limit?
YES
NO
Message
NO SAMPLE?
2
NO
YES
Wash cycle
Restart wash
interval timer
NOTES
1. Remote wash is triggered at
the closing of the switch. If
the switch is held closed,
only one wash cycle is
carried out.
2. The wash is inhibited if
there is no sample, no
sensor or the sensor cannot
measure correctly.
Figure 6.15
Wash logic