PR-23 instruction manual

PROCESS REFRACTOMETER PR-23

Sensor verification

Page loaded on February 2, 2009 at 03:19:28 PM

Sensor information

Serial number:

R05676

Test results

Nominal nD

Measurement

Sensor

at 25°C

at

actual
temp.

Measured

nD

Accuracy Result

Head

temp.
°C

Head

hum.
%

Supply

current
mA

1 1.3300

1.3297 1.3296

0.0001

pass

30

3

39

2 1.3700

1.3697 1.3696

0.0001

pass

30

3

39

3 1.4200

1.4196 1.4197

0.0000

pass

30

3

39

4 1.4700

1.4697 1.4697

0.0000

pass

30

3

39

5 1.5200

1.5196 1.5196

0.0000

pass

30

3

39

Verification result: pass

Traceability and conformance

Measurements should be done by using Cargille Refractive Index Liquids with

stated in accuracy of ±0.0002 R.I.
Cargille Index of Refraction Liquids are manufactured and calibrated using
instruments whose accuracy is verified by daily comparision to N.I.S.T. (N.B.S.)

traceable standards. N.I.S.T. standards are based on angle measurements using
a Wild divided circle spectrometers.

K-Patents PR-23 Refractometer System: Sensor verification

http://192.168.23.250/verif.shtml

1 of 1

2.2.2009 15:21

Figure 13.8

Instrument

verification certificate

13.4 Corrective action

If you get the message

VERIFICATION FAILED

(Figure 13.9), first check that the prism and

the sample holder are absolutely clean and the sample holder sits tightly on the
sensor tip before a standard liquid is applied. Make sure the standard liquids are in
good condition and not past their expiration date. Also, inspect the prism surface,
checking that it is smooth and glossy without any scratches. Then go back to the
Verification step 2 by pressing soft key

REPEAT

and repeat the whole verification

procedure.
You may want to double-check the following, as they are the most common reasons
for a failure of the verification: