beautypg.com

Side view top view, 1 welding isolation valve to pipe – K-Patents PR-23 User Manual

Page 156

background image

148

PR-23 instruction manual

The Safe-Drive™ system is mounted on a vertical pipe or horizontal pipe. When choos-
ing mounting location, keep in mind that you have to be able to lift the Retractor with
sensor inside over and off the isolation valve for sensor insertion and removal.

11.6

295

2

4

9

5

.0

1

272

10.71

1

2

7

1.27

32

0

.9

9

2

5

Side view

Top view

26.09.2013 IA

C

2929

1:5

SDI2 vertical pipe, space requirement

MTG

-

31.01.2014 AHä

General tolerances

Revision Scale

Appr.

Drawn

Drawing number

Drawing description

Mass

Material

Floor or
working platform

5

0

0

-

1

0

0

0

[

2

-

3

f

t]

1000 - 1500 [3 - 6 ft]

Figure 11.4

Mounting on a vertical pipe

B

11.60 295

8.86

225

Side view

Top view

1.34

34

5.01 127

F

10.78 274

1.93

49

Related drawing.

No modification permitted

without the approval of

the authorized person.

500 - 1000 [2 - 3 ft]

1000 - 1500 [3 - 6 ft]

Floor or

working platform

Appr.

Drawn

Drawing number

Drawing description

Mass

General tolerances

MTG

27.09.2013 IA

-

Revision

2931

A

Scale

1:5

SDI2 horizontal pipe, space requirement

Material

31.01.2014 AHä

Figure 11.5

Mounting on a horizontal pipe

11.4.1 Welding isolation valve to pipe

For the Safe-Drive™ Isolation valve, two holes – 50 mm (2") and 25 mm (1") – are
drilled to the pipe and the bridge between the holes is then removed. To help you place