3 specifications – K-Patents PR-23 User Manual
Page 148

140
PR-23 instruction manual
10.3 Specifications
10.3.1 Indicating transmitter specifications
Display:
320x240 pixel graphical LCD with LED backlight
Keypad:
18 membrane keys
Current output:
Two independent current sources, 4-20 mA, max.
load 1000 Ohm, galvanic isolation 1500 VDC or AC
(peak), hold function during prism wash
Power:
AC input 100-240 VAC, ±10%, 50/60 Hz/30 VA,
optional 24 VDC, Overvoltage Category II, Pollution
Degree 2 (DTR), Pollution Degree 3 (sensor)
Alarms/Wash relays:
Two built-in signal relays, max. 250 V/3 A
Input switches:
Four switch inputs.
Current outputs:
Two current outputs configurable independently to
indicate process concentration or temperature of
either sensor.
Sensor connectivity, DTR:
One or two sensors can be connected to the
transmitter. Sensors independent of each other:
own parameter sets and usable in different
applications.
Sensor connectivity, STR:
Only one sensor can be connected to the transmitter.
Used with the Intrinsically safe sensor PR-23-...-IA.
Transmitter protection class:
Enclosure IP66, Nema 4X
Installation:
Transmitter indoor, sensor indoor/outdoor, relative
humidity 80% max
Indicating transmitter weight:
4.5 kg (10 lbs)
10.3.2 Interconnecting cable specifications
Cable:
IEC 61158-2 compliant two-wire cable:
two signal wires and shield
copper area 0.8 mm
2
(18 AWG)
cable resistance 24 Ohm/km (per wire)
cable attenuation 3.0 dB/km @ 28 kHz
Cable length:
Standard 10 m (33 ft), max. total length 200 m
(660 ft)
Note:
For information on the intrinsically safe cabling for PR-23-...-IA, see Sec-
tion 9.11.2 on page 132.