4 prism wash systems, 1 prism coating, 2 prism wash – K-Patents PR-23 User Manual
Page 25
4 Prism wash systems
17
4 Prism wash systems
4.1 Prism coating
Deposit build-up on the prism surface disturbs the measurement. Look out for an
abnormally high concentration reading or an upward concentration (CONC) drift.
In most applications the prism will keep clean due to the self-cleaning effect. If coating
occurs, check the following:
− Sufficient flow velocity, see Section 2.2.3, “Check list for pipe mounting”.
− A temperature difference between process fluid and sensor probe may cause coat-
ing. This may happen with small flows if the thermal insulation is inadequate. In
some cases it helps to also insulate the clamp connector.
In case of a coating problem, the preferred solution is to try to increase the flow ve-
locity, e.g. by installing a pipe portion with smaller diameter.
Installing a wash nozzle can be considered, if increasing the velocity or using a flow
booster does not provide a solution (Section 4.2).
4.2 Prism wash
Three alternative media can be used for prism wash: steam, water, high pressure wa-
ter. The built-in relays of the Indicating transmitter can be configured to control the
prism wash cycle, see Section 6.3, “Configuring relays”.
Important:
In food industry applications, wash media must be culinary steam or
safe water. Shut-off valve and check valve must meet 3-A sanitary standards (see Fig-
ures 4.2 and 4.5.
Culinary Steam: Shall mean steam produced using a system meeting criteria in the 3-A
Accepted Practices for a Method of Producing Steam of a Culinary Quality, Number
609.
Safe Water: Shall mean water from a supply properly located, protected, and operated,
and shall be of safe, sanitary quality. The water shall meet the standards prescribed
in the National Primary Drinking Water Regulation of the Environmental Protection
Agency (EPA) as referenced in The Code of Federal Regulations (CFR), Title 40, Parts
141, 142, and 143.