PR-23 instruction manual

8.2 Measurement

8.2.1 Message

OUTSIDE LIGHT ERROR

Cause: The measurement is not possible because too much outside light reaches the
camera.

Action: Identify the light source (for example sun shining into an open tank or a
translucent pipe) and block the light from getting to the prism at the sensor tip.

8.2.2 Message

NO OPTICAL IMAGE

The optical image can be seen from selecting

3 SENSOR STATUS

at the Main menu, Sec-

tion 5.4.1. There are several possible causes:

1. The prism is heavily coated, Section 4.1. Perform prism wash if available, Sec-

tion 5.1.3 “Testing prism wash”. Remove sensor from line and clean prism manu-
ally.

2. There is moisture condensation in the sensor head, see Section 8.1.7.
3. The sensor head temperature is too high, see Section 8.1.8.
4. The light source is faulty. When the sensor is removed from the process, the yellow

flashing light can be seen through the prism. Note: The light is only visible at an
oblique angle. Also check the LED value in the Sensor status display (select

3 Sensor

status

in the Main menu); if the value is clearly below 100, LED fault is not likely.

5. There are negative spikes in the optical image. The probable cause is dust or fin-

gerprints on the CCD window.

6. The CCD card in the sensor is faulty. See Section 7.3, “Disassembling and assem-

bling the sensor”.

8.2.3 Message

PRISM COATED

Cause: The optical surface of the prism is coated by the process medium or impurities
in the process medium.

Action: Perform prism wash if available, Section 5.1.3 “Testing prism wash”. Remove
sensor from line and clean prism manually.

If the problem is recurrent, consider improving the flow conditions (see Section 2.2,
“Mounting the sensor”) or, if prism wash is available, adjust the wash parameters, see
Section 6.7, “Configuring prism wash”.