9.9.4 PR-23-W mounting specifics
PR-23-W Saunders valve body flowcell can be mounted either vertically or horizon-
tally. Special sensor support is not needed as the valve body (piping) supports the
sensor. Either way sensor cover should always be horizontal to avoid sedimentation
or gas/air pocket on the prism. Also installation after pump, before valve and low
installation point will reduce risk of air/gas pocket. Recommended flow velocity is
1.5–6m/s (5–20ft/s).