beautypg.com

K-Patents PR-23 User Manual

Page 43

background image

5 Startup and use

35

Empty pipe

Normal conditions

Figure 5.8

Typical optical images with IDS

Figure 5.9

A slope graph with IDS

It should be noted that the "empty" optical image may have vertical left and/or right
edge close to the edge of the image. In the example, only the right edge is visible.

5.4.3 Diagnostic values

The values at the left of the graph are used for diagnostic purposes:

CONC

is the final concentration value including Field calibration adjustment, see

Figure 6.12.

TEMP

, see Section 5.4.4.

CCD

gives the position of the shadow edge on CCD in %.

nD

is the refractive index value n

D

from the sensor.

CALC

is the calculated concentration value without Field calibration adjustment,

Section 6.6.3

QF

or Quality Factor is a value in the range 0–200. It measures the image sharpness,

a typical good value is 100. A QF value below 40 usually indicates prism coating.

LED

is a measure of the amount of light from the light source in %. Should be below

100 %.

HD TMP

= sensor head temperature, see Section 5.4.4.

HD HUM

= sensor head humidity, see Section 5.4.5.

I_SNS

value shows the current to sensor, the nominal value is 40 mA.

DTR TMP

= Indicating transmitter temperature, see Section 5.4.4.