30

PR-23 instruction manual

6. For a connected sensor, the diagnostic message at start-up should be

NORMAL OPERA-

TION

or, if the process pipe is empty,

NO SAMPLE

. Otherwise, see Section 8.4, “Diagnos-

tic messages table”.

7. The TEMP value should show the current process temperature.

8. The value and the correct setup of the two mA output signals can be checked by

selecting

DESCRIPTION

in the Main menu and then

mA OUTPUTS

in the Description menu

(Section 5.3).

9. If internal relays or switch inputs are used, their settings can also be checked

through the Description menu (Section 5.3).

5.1.2 Calibration check

Wait until normal process conditions occur. The concentration reading is precali-
brated at delivery and a copy of the Sensor calibration certificate is inside the Indi-
cating transmitter. If the diagnostic message is

NORMAL OPERATION

but the concentration

reading does not agree with the laboratory results, then consult Section 6.6, “Calibrat-
ing the concentration measurement”.

5.1.3 Testing prism wash

1. Check that the steam or water washing parts are connected (Section 4.2.2, “Prism

wash systems”).

2. In the Main display, press

MENU

. Then press 3 (to give the command

SENSOR STATUS

).

In this Sensor status display by pressing the soft key

WASH

. If soft key

WASH

does not

appear, no internal relay is configured for this purpose.

3. Check the n

D

reading, for a successful wash it must drop below 1.34 during steam

wash and drop to approximately 1.33 during water wash.

Important: Before testing prism wash, check that there is liquid in the pipe in front
of the refractometer sensor.

5.2 Using the Indicating transmitter

The Indicating transmitter DTR receives the refractive index value n

D

and the process

temperature from the sensor(s). Starting from these values, it calculates the concen-
tration of the process media for display and further transmission. The DTR can also
be programmed to give alarm for high or low concentration. If the refractometer has
a prism wash system, the DTR can control the wash with its built-in timer.

For information on how to use the Indicating transmitter DTR for configuration and
calibration, see Chapter 6, “Configuration and calibration”.