6 Configuration and calibration
47
NAMUR NE 43 uses the 3.8 to 20.5 mA signal range for measurement information,
with ≥21 mA or ≤3.6 mA to indicate diagnostic failures (see figure 6.11). With that
information, it is easier to detect a failure condition on a refractometer, for example,
it clearly tells you whether you have an empty pipe or a failed instrument.
0
3.6
3.8
4
20
20.5
21
mA
failure
OK
Measurement Data
failure
Figure 6.11
Default mA output values
−
6 MANUAL SET
allows you to set different output values to check the output signal.
Press the
BACK
key to return normal output function.
Note: If you want to ’turn off’ the mA output, select
NOT DEFINED
in the Source menu.