6 Configuration and calibration

47

NAMUR NE 43 uses the 3.8 to 20.5 mA signal range for measurement information,
with ≥21 mA or ≤3.6 mA to indicate diagnostic failures (see figure 6.11). With that
information, it is easier to detect a failure condition on a refractometer, for example,
it clearly tells you whether you have an empty pipe or a failed instrument.

0

3.6

3.8

4

20

20.5

21

mA

failure

OK

Measurement Data

failure

Figure 6.11

Default mA output values

6 MANUAL SET

allows you to set different output values to check the output signal.

Press the

BACK

key to return normal output function.

Note: If you want to ’turn off’ the mA output, select

NOT DEFINED

in the Source menu.