beautypg.com

2 specifications – K-Patents PR-23 User Manual

Page 152

background image

144

PR-23 instruction manual

11.2 Specifications

Refractive Index range:

Full range n

D

1.3200–1.5300 (corresponds

to hot water – 100 Brix)

Accuracy:

Refractive index n

D

± 0.0002 (corresponds

typically to ± 0.1 % by weight)
Repeatability and stability correspond to
accuracy

Speed of response:

1 s undamped, damping time selectable up
to 5 min

Calibration:

With Cargille certified refractive index liq-
uids over full range of n

D

1.3200–1.5300

CORE-Optics:

No mechanical adjustments (US Patent No.
US6067151)

Digital measurement:

3648 pixel CCD element

Light source:

Light emitting diode (LED) 589 nm wave-
length, sodium light

Temperature sensor:

Built-in Pt-1000

Temperature compensation:

Automatic, digital compensation

Instrument verification:

With certified refractive index liquids and
K-Patents documented procedure

Ambient temperature:

Sensor: max. 45 °C (113 °F), min. -20 °C
(-4 °F); Indicating transmitter: max. 50 °C
(122 °F), min. 0 °C (32 °F)

SAFE-DRIVE™ SENSOR PR-23-SD

(Patent Pending)

AND ISOLATION VALVE SDI-23
Isolation valve connection:

Safe-Drive™ flange DN 40 PN25
(Patent pending)

Process pressure:

Static pressure up to 20 bar (300 psi),
operational pressure up to 10 bar (150 psi)

Process temperature:

-20 °C–170 °C (-4 °F–340 °F)

Sensor process wetted parts,

SAF 2205, Duplex steel SS 2377, Werkstoff-
Nr.

standard:

1.4462, UNS S31803, prism spinel, prism
gaskets MTF (Modified Teflon)

Sensor protection class:

IP67, Nema 4X

Isolation valve

SAF 2205, Duplex steel SS 2377, Werkstoff-
Nr.

process wetted parts:

1.4462, UNS S31803, AISI 316 L, flange
gasket Viton

®

, lip seals Bronze Teflon

®

and ELGILOY, AISI 301 spring

Isolation valve

By welding to pipe sizes of 2"–24", for both

process connection:

vertical and horizontal pipe lines

Prism wash:

Retractable steam wash nozzle with check
valves

Sensor and valve weight:

10.5 kg (23 lbs)