4 configuring input switches – K-Patents PR-23 User Manual
Page 52
44
PR-23 instruction manual
6.4 Configuring input switches
For the electrical properties of the four input switches, see Section 3.3. To see which
switches are closed, check the Description menu, see Section 5.3. To configure the
switches, follow the instructions below:
1. Select Menu to get to the Main menu.
2. Select
5 CALIBRATION
from the Main menu.
3. Select
5 SWITCHES
from the Calibration menu.
4. Select the switch, 1, 2, 3 or 4, to be configured. You now get the Switch menu, see
Figure 6.8 below.
Figure 6.8
Switch menu
5. First select
1 SENSOR
to assign the chosen switch to a given sensor.
Note: The selection line will automatically go to the currently valid setting, i.e. in
Figure 6.9 next page Switch 1 has been assigned to Sensor A.
Figure 6.9
Arrival to Switch 1’s sensor selection menu, sensor A selected
6. In the Switch menu, select
2 FUNCTION
to set switch function.
Note: The current assignment of the switch is shown at the bottom of the Switch
menu display, e.g. Figure 6.9 Switch 1 is assigned to Sensor A with function Wash
stop.