158
PR-23 instruction manual
9
1 Turn the hand-wheel left to drop the thread.
2 Lift off the Outer casing.
10
1 Lift up the latch of the Inner casing to unlock.
2 Turn the casing 60 degrees left to release it from
the flange.
11
Lift the Inner casing away from the sensor head.
12
Power off the DTR. Connect sensor cable to the
DTR.
1 Take off sensor nameplate and the gasket
underneath.
2 Put interconnecting cable through the cable
gland.
3 Connect the interconnecting cable to the sensor.
4 Screw the cable gland to the sensor.
5 Fit the gasket and nameplate onto the sensor
and screw the nameplate back on.
Turn on DTR power to power up the Safe-Drive™
system.