Guralp Systems CMG-DCM build <10,000 User Manual
Page 132

Acquisition Modules and Platinum Firmware
Receiving Data
10.2.1 Configurable parameters in simple mode
The configurable parameters for GCF BRP receiver are contained in a single
form in simple mode:
User description: Sets the name of the service; this should be set to a
meaningful name for the data that it will be receiving, such as the IP or
hostname of the network digitiser.
User label: Identify the particular client instance in log-files (optional).
Enable: Causes this service to start automatically when the system is
re-booted. If this check-box is cleared, the service will need to be started
manually (from the “Control” → “Services” menu)
Delete: Causes the configuration for this instance to be removed from the
system when the form is submitted.
Remote Server: Specify the hostname or IP address of the network digitiser
Remote service: Specify the port (name or number) that the digitiser is
transmitting on.
Allow disconnects: If checked, the instance will attempt to automatically
recover from lost connections by trying to reconnect to the server.
Disable rewind: If checked no attempts will be made to request missing data
blocks. This should only be selected if the server is unable to fulfil such
requests.
132