Appendix d profinia, System error messages – Bio-Rad Profinia™ Protein Purification Instrument User Manual
Page 183

Appendix D
Profinia
™
System Error Messages
The error messages listed below are displayed on the Profinia touch screen when instrument
operation errors occur.
Message
Possible Cause(s) and Solutions
Cannot Talk to I2C device X*
Interprocessor communication problems;
possibly due to loose connections.
Check internal connections and try again.
Valve configuration error
Dirty valves, defective valves, or defective
cables.
Bad valve X*
Position error: valve X*
Can’t home: valve X*
Can’t find prop valve
Internal error: X*
Power cycle the unit; contact Bio-Rad if error
persists.
Air in line detected!
Replace or add reagent.
Check all reagents and deionized water.
Press Start to re-prime the line and continue
the method.
Press Stop to terminate run.
Pump Error
Contact Bio-Rad if error persists.
Over pressure detected @60 psi
Correct and continue.
operations suspended
Over pressure detected-paused
Correct and continue
check cartridges and method
Over pressure-did you remove
Correct and cycle power
the waste line shipping plug?
Method storage is full saving
Delete a stored method or proceed without
(35 methods can be stored; user can
continue without saving methods).
Internal hardware error, disconnected
Check connections; contact Bio-Rad if error
components, check device X*
persists.
Trouble with PV board, error X*
Check connections.
Valve move while pumping, run paused
Contact Bio-Rad if error persists.
There is a problem using your memory
If problem persists, power-cycle the unit and
stick. Please re-insert it and try again.
retry. Use a different portable memory
device.