beautypg.com

Power, Internal power – ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual

Page 47

background image

2-10 ADS TRITON+ Manual

Power

Internal Power

Power is supplied to the TRITON+ flow monitor by an internal 12-
volt IS battery pack (ADS p/n 8000-0043-02/04). The battery pack
resides within the TRITON+ enclosure and powers monitor
operations. A battery fuse board is attached to the battery pack to
achieve intrinsic safety.

TRITON+ 12-volt battery

The power regulator, built into the dome assembly that covers and
protects the processor board, is critical to intrinsically safe
operation. It regulates the voltage supplied from the internal IS
battery pack to the TRITON+, ensuring that the monitor always
operates within its IS certification requirements. It also includes
replaceable fuses and current-limiting components that will limit
and/or interrupt power if a fault condition occurs.