ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual
Page 205

6-18 ADS TRITON+ Manual
Modify the parameters as necessary, and then select OK once
complete.
DAC Gain Choose the method for managing the
amplification of the analog signal that returns to the sensor:
− Auto Select this option to allow the monitor to
automatically optimize the amplification of the signal.
− Manual Select this option to enter a fixed value
between -12db (lower amplification) and +12db (high
amplification) at which to amplify the signal.
Ping Amplitude Choose the method for managing the
amplification of the analog signal that is sent out, or
transmitted from, the sensor. Smaller pipes typically
require weaker (quieter) amplification; larger pipes
typically require stronger (louder) amplification.
− Auto Select this option to allow the monitor to
automatically optimize the amplification of the signal.
− Manual Select this option to enter the percentage of
full amplification at which to transmit the signal.
Enter a percentage between 10 and 100 percent, where
100 percent represents maximum amplification.