6-30 ADS TRITON+ Manual

Mode Select the mode in which you want the smart depth
sensor to operate based on environmental factors or site
constraints. ADS recommends using the Normal setting for
pipe heights ranging from 6 to 24 inches (15 to 61 cm) and
High for pipe heights greater than 24 inches (61 cm) under
most conditions. However, if erratic Smart Depth readings
occur, consider changing the Power setting.

Blanking Select the distance from the horizontal face of the
Surface Combo Sensor within which to ignore data potentially
representing erroneous readings. Select the Auto option
(default) to ensure the monitor disregards depths occurring
within a distance from the horizontal face of the sensor
equivalent to 5 percent of the range from the sensor face to the
bottom of the pipe. This option typically represents the vast
majority of pipe/flow conditions; therefore, ADS strongly
recommends leaving this option selected for most applications.
Choose one of the other options only when spikes in the data
are occurring beyond 5 percent of this range and under direct
consultation with an ADS representative.

Store Data Select this checkbox to ensure the monitor logs
the smart depth data to memory. This option is selected by
default.

Advanced Select this button to access the Advanced
Parameters
dialog.