Maintenance and Troubleshooting 10-51

Problem

Velocity data does not fluctuate much (but is not 0).

Possible

Causes

Sensor may be dirty.

Minimum flows may be occurring within the sensor
deadband (extending 1.0 inches (25 mm) up from the
bottom of the Peak Combo Sensor).

Velocity function may be broken.

Possible

Solutions

Clean the sensor’s transmit/receive surfaces.

Verify that minimum flows are occurring outside the
sensor deadband.

Check the sensor connections.

Check the velocity parameters.

Contact your regional ADS representative.

Replace the velocity sensor.