Configuration and Activation 6-79

Performing Confirmations

Confirmations allow you to compare measurements taken by the
sensors to measurements taken manually in the field. These
comparisons help to verify that the sensors are functioning properly
and determine whether specific device parameters may require
modification.

Note: Confirmations for TRITON+ monitors equipped
with SIMs with dynamic IP addresses must be performed
serially.

Confirm the sensors in the following way:

1. Select the monitor for which you want to perform

confirmations from the Location Name field in the Monitor
section on the ADS Qstart dialog.

2. Select the device with the sensor(s) you want to confirm from

the Devices list.

3. Click on the Confirm button on the Monitoring Point tab.

Qstart establishes a connection with the monitor and displays
the Confirm Device dialog.

4. Select the point of reference from which to compare manual

depth to the actual sensor readings from the drop-down list in
the Field Measurement section.

DOF Represents the depth of the flow from the bottom of

the pipe to the flow surface.

Air DOF Crown Represents the distance (or range) from

the top (or crown) of the pipe to the flow surface.

Air DOF Bat Represents the distance (or range) from the

ultrasonic depth sensor face (crystal surface) to the flow
surface.