Performing confirmations – ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual
Page 266

Configuration and Activation 6-79
Performing Confirmations
Confirmations allow you to compare measurements taken by the
sensors to measurements taken manually in the field. These
comparisons help to verify that the sensors are functioning properly
and determine whether specific device parameters may require
modification.
Note: Confirmations for TRITON+ monitors equipped
with SIMs with dynamic IP addresses must be performed
serially.
Confirm the sensors in the following way:
1. Select the monitor for which you want to perform
confirmations from the Location Name field in the Monitor
section on the ADS Qstart dialog.
2. Select the device with the sensor(s) you want to confirm from
the Devices list.
3. Click on the Confirm button on the Monitoring Point tab.
Qstart establishes a connection with the monitor and displays
the Confirm Device dialog.
4. Select the point of reference from which to compare manual
depth to the actual sensor readings from the drop-down list in
the Field Measurement section.
DOF Represents the depth of the flow from the bottom of
the pipe to the flow surface.
Air DOF Crown Represents the distance (or range) from
the top (or crown) of the pipe to the flow surface.
Air DOF Bat Represents the distance (or range) from the
ultrasonic depth sensor face (crystal surface) to the flow
surface.