beautypg.com

ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual

Page 106

background image

Sensor Installation and Connection

3-45

3. (applies only when mounting the sensor no more than 15

degrees (6:30 position) up the side of the pipe) Mount the
sensor onto the inside of the band with two M3 x 10-mm
countersink screws, making sure the blunt end (i.e., nose) of the
sensor is facing the upstream edge of the ring (edge opposite
cable tie holes). If mounting holes do not exist on the band at
or near the desired mounting location for the sensor, drill
holes in the band using a 7/32-inch (3-mm) drill bit.

Attaching Peak Combo Sensor to band

4. (applies only when mounting the sensor between 15 and 45

degrees up the side of the pipe) Use the silt mount adapter
(ADS p/n 8000-0271) to attach the sensor to the band in the
following way:

 Seat the Peak Combo Sensor onto the slightly elevated

portion of the top of the silt mount adapter. The elevated
portion of the adapter should insert easily into the inset
portion of the bottom of the sensor. Orient the sensor so
that the sensor cable is exiting the sharp, or wedged, side
of the adapter.

 From the bottom of the adapter, use two M3 x 10mm

stainless steel flathead screws (do not substitute any other
screws
) to secure the sensor to the adapter. The predrilled
screw holes in the adapter and in the bottom of the sensor
should be in alignment if the sensor is seated properly on
the adapter.