ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual
Page 295

Analog and Digital I/O 8-9
with the analog output parameters that indicate the values or
measurements that correspond to the analog signals within the
4-20mA range. Refer to Chapter 6, Configuration and
Activation, for instructions on supplying this information in
Qstart.
3. Configure the device or instrument receiving the analog signals
with the range and engineering units corresponding to the 4-
20mA from the TRITON+.
4. Connect the analog output wires to the ADS XIO in the
following way:
Note: Make sure the power is turned off to the third-party
device and XIO.
Release the latches and open the clear, front door on the
XIO.
Run the wires up through the left cable gland on the XIO
into the enclosure. ADS recommends using 18-gauge
wiring. You may need to loosen the plastic nut on the
cable gland on the outside of the XIO to feed the wires
through.