Configuration and Activation 6-81

If a difference exists between the manual velocity
measurements and the velocity sensor measurements, you can
compensate for this difference by adjusting certain velocity
sensor parameters, particularly Maximum Velocity and
Sensitivity.

Note: If you change a device parameter, such as an
electronic offset, you must reactivate the monitor.

13. (optional) Click on the Read button again to take additional

measurements, if necessary.

14. Click on the Close button once you are finished performing

confirmations for the selected device to exit the Confirm
Device
dialog.

15. Repeat steps 2 through 12 for each additional device that

includes a sensor(s) you want to confirm.

Note: If the location supports a second monitoring point,
remember to confirm the devices assigned to that
monitoring point, as necessary.

16. If a change has been made to any sensor parameter, reactivate

the monitor. See page 6-72 for more information on
reactivating a monitor. If you didn’t change any device
parameters, there is no need to reactivate the monitor.

17. Click on the green Hangup button in the Functions section to

disconnect from the monitor.

Qstart generates a log entry of the confirmation activity, which
includes the field and monitor readings, the name of the crew,
and any associated comments, and saves it to the local
directory or network. It also generates a log entry for each
user-initiated reading taken by the monitor that includes the
sensor fired, date and time at which the measurement was
taken, current device parameters corresponding to the sensor,
and reading and associated quality. These entries are
available for viewing through the log viewer in Qstart.