ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual
Page 405

D-1
A P P E N D I X D
System Configuration and Setup to
Support the Telog Ru-33
Using a battery-powered ADS
®
TRITON+
™
to support the Telog
®
Ru-33 Recording Telemetry Unit (RTU) requires some limited
configuration and setup. These procedures involve designating the
appropriate identification information to enable recognition and
data exchange between the ADS and Telog units and using an
interface cable to connect the monitor to the RTU.
Note: The Telog Ru-33 must have firmware version
3.300 or later to effectively request data from the
TRITON+ monitor.
Connecting the units together primarily involves running the ADS
Triton-Telog Comm Cable from the battery-powered Ru-33 to the
battery-powered TRITON+ monitor. This application requires
installing both units in the manhole, a hazardous environment.
However, ADS offers an optional extension cable (ADS p/n 8000-
0044-25) that allows you to install the Telog unit outside the
hazardous environment.
Note: The Telog Ru-33 does not possess IS (intrinsically-
safe) certification. Therefore, an installation involving
connecting the Telog to the IS TRITON+ with both units
residing inside the manhole will not be considered
intrinsically safe. The installation will be considered