Sensor Installation and Connection

3-69

6. (applicable only when performing water quality sampling)

Connect the appropriate sampler cable to the COMM + EXT
PWR
port. Tighten the rotating connector collar in a
clockwise direction until it clicks, and verify that it is seated
correctly.

7. For locations exhibiting the presence of hydrogen sulfide

and/or high moisture, provide added protection by wrapping
the connection with mastic tape.

8. Seal any unused connectors with protective caps. If

necessary, obtain additional caps from ADS.

Note: An air valve is present in the monitor top. This
valve exists only for testing purposes during the
manufacturing process. Do not attempt to fill the monitor
with air or pressurize the monitor!

Securing the Dryer Tube to the Monitor

The Peak Combo Sensor and Surface Combo Sensor each contain a
pressure depth sensor with an associated dryer tube. For each
combo sensor connected to the monitor, you must secure the
corresponding dryer tube (attached to a cable running from the
sensor connector) to the monitor in the following way:

1. Attach the dryer tube from the pressure depth sensor in the

vertical position to the monitor handle using cable ties. Make
sure the clear, flexible tubing coming from the sensor connector
to the top of the dryer tube is not kinked. If necessary, use a
loose cable tie to hold it in an unkinked position.