ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual
Page 264

Configuration and Activation 6-77
The monitor takes the sensor measurements and displays the
readings in the Results window. The results may include the
specific reading (e.g., depth or velocity), the quality of the
reading, the maximum signal strength, and the temperature.
When taking depth measurements, the dialog also displays a
cross section of the pipe, offering a visual perspective of the
depth of the flow.
Note: Pressure depth readings compensate for the
physical and electronic offsets, and the Automatic Pressure
Calibration value (PRESSK and PRESSK2).
8. Verify the accuracy, consistency, and quality of the readings,
and then edit the corresponding device parameters as necessary.
9. (optional) Repeat steps 3 through 10 for every other sensor on
the device dialog on which you want to run diagnostics.
10. Click on the OK button once you are finished running
diagnostics for all the appropriate sensors corresponding to the
current device. This saves any changes made to the parameters
to the local directory or network. Clicking on the Cancel
button closes the device dialog without saving any parameter
modifications.
11. If any changes have been made to any of the device parameters,
you must reactivate the monitor by clicking on the Activate
button in the Functions section.
Note: There is no need to reactivate a monitor if there
were no parameter changes!
12. Once activation is complete, the status bar displays Activate
successful.
13. Click on the Hangup button in the Functions section when you
are finished running diagnostics on sensors associated with the
current monitor.
14. Click on the Save button to save any changes in the device
parameters to the local directory or network.