ADS Environmental Services ADS TRITON+ QR 775027 A3 User Manual
Page 379

10-58 ADS TRITON+ Manual
Problem
Data or readings received at customer input device from
XIO 4-20 mA output are not accurate or were
unexpected.
Possible
Causes
Wiring from XIO output to external 4-20 mA input device
may not be properly connected. (Note: The XIO
provides loop power to both 4-20 mA outputs).
XIO Device may not have been properly configured
through Qstart.
Analog module in XIO may not have been setup
properly or may be defective.
Possible
Solutions
Verify wiring between XIO output and the external input
device has been properly connected.
Verify that the correct entity has been selected and that
the scale has been set properly in Qstart. If necessary,
adjust the configuration and reactivate the monitor.
Contact your regional ADS representative.