4. Detailed description
711 Liquino, Instructions for use
81
4.4.2
Examples of reports - XDOS mode
Temperature-controlled dosing with volume and time selection
Parameter Report
711 Liquino 0220 711.0011
date 1998-09-30 time 11:51
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: XDOS user method: TempDos1
lot no. 12-2
parameters
>XDOS parameter
XDOS type: volume&time
volume 10.0 mL
dosing time 00:02:00
temp. measuring: on
rec. interval 00:00:10
fill rate max mL/min
auto fill: on
>XDOS dosing units
setup: single
dosing drive: DOS1
dos.unit: D3
port use: custom
dosing port 1
filling port 2
waste port 1
>XDOS monitoring
temperature: on
low limit 20.0 °C
upper limit 30.0 °C
action: wait
alarm pins: none
volume: off
>XDOS preselections
req.ident: id1
>>XDOS prompts
id1 lot no.
id2 id2
id3 id3
start delay 00:00:00
>XDOS reports
temp. auto scale: on
report1: result
report2: curve
report3: comb. list
report4: off
report5: off
report6: off
report7: off
report8: off
------------
← instrument number and program version
← date and time
← 3-line report header
← mode and method name
← ID1, identification with preselected display text
← dosing parameters
← dosing unit settings
← monitoring function settings
← request settings
← report definitions
← identification of reproduced reports
Result Report
711 Liquino 0220 711.0011
date 1998-09-30 time 11:39
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: XDOS user method: TempDos1
lot no. 12-2
total dosing time = 000:02:17
total volume = 10.000 mL
dosing rate = auto
============
← see above
← effective time *
← dosed volume
← dosing rate
← identification of original report
*
total time
including waiting times during infringements of temperature limits or during filling