3.2 Tutorial
711 Liquino, Instructions for use
48
2x <>
dos. uni t : st andar d
•
In this menu item the 711 Liquino is informed about the tubing of the
Dosino. Press
•
Skip the next menu item with
Di spl ay:
par amet er s
>XDOS moni t or i ng
•
As the submenu ‘>XDOS dosing units' does not contain any further
menu items, an automatic return is made to the submenu selection.
Press
Di spl ay:
>XDOS moni t or i ng
t emper at ur e: of f
•
Use the
temperature monitoring.
Di spl ay:
>XDOS moni t or i ng
l ow l i mi t - 70 ° C
•
The lower temperature limit is of no interest in this application. Accept
the default value with
Di spl ay:
>XDOS moni t or i ng
upper l i mi t 200 ° C
•
Set the upper temperature limit to 10 °C. Confirm the entry with
Di spl ay:
>XDOS moni t or i ng
act i on: none
•
The action to be taken if the temperature exceeds the set limits is de-
termined here. Use
•
No further entries are necessary in the ‘XDOS monitoring’ submenu.
Exit with the
Di spl ay:
par amet er s
XDOS moni t or i ng
•
Press <> 2x to switch to the submenu ‘XDOS reports’.
Di spl ay:
par amet er s
>XDOS r epor t s
•
Press
Di spl ay:
>XDOS r epor t s
t emp. aut o scal e: on
*) If no printer or PC is connected the parameter menu can be exited with
at "Running the method“.