Metrohm 711 Liquino User Manual
Page 85
![background image](/manuals/587118/85/background.png)
4. Detailed description
711 Liquino, Instructions for use
79
par amet er s
>XDOS moni t or i ng
Monitoring function settings
Navigate with
>XDOS moni t or i ng
t emper at ur e: of f
Temperature monitoring
on, off
>XDOS moni t or i ng
l ow l i mi t - 70 ° C
Lower temperature limit
-70…200.0
>XDOS moni t or i ng
upper l i mi t 200 ° C
Upper temperature limit
-70…200.0
>XDOS moni t or i ng
act i on: none
Reaction if upper or lower temperature limit is
infringed
none,
stop,
hold
wait
= no reaction (alarm pins still active)
= dosing is stopped
= dosing is interrupted and can be continued with
= dosing is interrupted and automatically continued
as soon as the measuring temperature is again within
the limits.
>XDOS moni t or i ng
al ar m pi ns: none
Definition of the active alarm pins of the re-
mote connection
none,
under,
over
all
= no activation
= Output2 lead (Lower Limit, pin 4) activated when tem-
perature too low
= Output1 lead (Upper Limit, pin 18) activated when
temperature too high
= Output1 and Output2 activated when corresponding
temperature limits are infringed (see above)
>XDOS moni t or i ng
vol ume: of f
Volume monitoring function
on, off
>XDOS moni t or i ng
cr i t . vol ume 10000 mL
Absolute volume limit for terminating dosing
0.0…99999.9
onl y f or ' vol ume: on'
par amet er s
>XDOS pr esel ect i ons
Definition of requests at start of dosing
Navigate with
>XDOS pr esel ect i ons
r eq. i dent : of f
Automatic request of up to
3 identification texts
id1,
id1&2,
all,
off
= only ID1 is requested
= ID1 and ID2 are requested
= ID1, ID2 and ID3 are requested
= no request
Use identification texts (ID1…3) for identifying your sample or your test. The
IDs will be printed out in the report of the dosing procedure so that the indi-
vidual result reports or curves are identified unambiguously.
The alarm pins can be
used to switch an external
instrument connected to
the remote socket, e.g. a
heating bath or a
thermostat.
You are recommended to
use Metrohm Relay Box
731.