Metrohm 711 Liquino User Manual
Page 120
![background image](/manuals/587118/120/background.png)
4.8 SMPL mode - fully automatic sampling
711 Liquino, Instructions for use
114
>SMPL sampl i ng oper at i on
sampl e r i nse r ep. 3
Number of times for pre-rinsing dosing
cylinder
1…10 recommended: 1 pre-rinse
>SMPL sampl i ng oper at i on
aspi r . r at e max mL/ mi n
Aspiration rate for sample transfer
0.001…166.66,
max
= highest possible aspiration rate (depends on type of
dosing unit), select with
>SMPL sampl i ng oper at i on
dosi ng r at e max mL/ mi n
Dosing rate for sample transfer
0.001…166.66,
max
= highest possible dosing rate (depends on type of
dosing unit), select with
>SMPL sampl i ng oper at i on
sol v. f i l l max mL/ mi n
Filling rate for dilution of sample
0.001…166.66,
max
= highest possible filling rate (depends on type of
dosing unit), select with
>SMPL sampl i ng oper at i on
ai r gap l engt h 10 mm
Length of separating bubble in aspiration tub-
ing
1…100
par amet er s
>SMPL meas. cel l r i nsi ng
Settings for rinsing the measuring cell
Navigate with
>SMPL meas. cel l r i nsi ng
ext er nal pump: of f
External pumps for aspirating and rinsing the
measuring cell
on, off
>SMPL meas. cel l r i nsi ng
r i nsi ng r epet i t i ons 3
Number of aspiration and rinsing processes
1…10 recommended: twice
>SMPL meas. cel l r i nsi ng
r i nse t i me 00: 01: 00
Rinsing time with external pump (solvent addi-
tion)
HH:MM:SS
only for 'external pump on'
correction of input with
>SMPL meas. cel l r i nsi ng
aspi r . t i me 00: 01: 00
Length of aspiration with external pump
HH:MM:SS
only for 'external pump on'
correction of input with
>SMPL meas. cel l r i nsi ng
aspi r . vol ume 10 mL
Aspiration volume of cell
0.1…500
with 'external pump off'
>SMPL meas. cel l r i nsi ng
aspi r . r at e max mL/ mi n
Aspiration rate for aspirating the cell
0.001…166.66,
max
with 'external pump off'
= highest possible aspiration rate (depends on type of
dosing unit), select with
Maximum values for
dosing and filling rates,
etc.
Cylinder vol. Max. rate
2 mL
6.66 mL/min
5 mL
16.66 mL/min
10 mL
33.33 mL/min
20 mL
66.66 mL/min
50 mL 166.66 mL/min
These are pure piston
stroke rates. Filling times
are not taken into account.