2 error messages – Metrohm 711 Liquino User Manual
Page 150
![background image](/manuals/587118/150/background.png)
6.2 Error messages
711 Liquino, Instructions for use
144
6.2
Error messages
* 50mL cyl . not possi bl e
In the PIP mode 50 mL dosing units cannot be used. Press
and
* * * bat t er y down
The memory card battery is empty. Insert the card in the instrument
and replace the battery according to the instructions given on the
leaflet accompanying the memory card. Press
ing the battery enter the date for the next battery change under
'
or
* * * bat t er y down
If no memory card is inserted and this message appears when the
Liquino is switched on then the support battery in the 711 Liquino is
empty. Please contact the Metrohm service department.
* * * change bat t er y
The date for the regular memory card battery change has been
reached. Change the battery according to the instructions given on
the leaflet accompanying the memory card. Leave the memory card
inserted in the instrument. Press
enter the date for the next battery change under '
'>change battery'.
* * * car d f ul l
No more free memory space is available on the memory card. Delete
methods which are no longer needed or use a different card. Press
* car d not accessi bl e
The memory card is being accessed simultaneously by the remote
control and by manual operation. Press
* * * car d r ead/ wr i t e er r or
The Liquino cannot read the memory card or cannot write on it.
Check whether the memory card has been inserted correctly. The
card may have been damaged by external influences. Press
* * * car d unf or mat ed
Before a memory card is used for the first time it must be formatted in
the Liquino. Press
* car d wr i t e- pr ot ect ed
The memory card is write-protected. Remove the write protection
(see leaflet accompanying memory card) or use a different card.
Press
* * * dat a l ost
The inserted memory card is damaged and can no longer be read.
Press
* * di r . al r eady exi st s
A directory with the given name already exists. Select another name
for the new directory. Press
DOSxx: f i l l f i r st !
The piston of Dosino xx is in an undefined position. Press
and then
DOSxx: i nval i d dos. por t
The multiple stopcock of Dosino xx is in an undefined position. Press
DOSxx: not r eady
Dosino xx is not connected or no dosing unit is attached. The ex-
change unit may not be mounted correctly. Correct the fault and
press
DOSxx: over l oad!
Dosino xx is blocked. A tubing or the multiple stopcock may be
blocked. Check the tubing and, if necessary, remove the dosing unit
(not the Dosino!). Then press