3 Introduction
711 Liquino, Instructions for use
47
PARAM
B
•
The key is used to access the parameter menu, where all
the possible settings for an XDOS method can be found.
Di spl ay:
par amet er s
>XDOS par amet er s
•
The submenu ‘XDOS parameters’ contains the main settings for an
XDOS method. Press
Di spl ay:
>XDOS par amet er s
XDOS Type: vol umen&r at e
•
Press
Di spl ay:
>XDOS par amet er s
vol ume 1. 000 mL
•
Enter the volume which is to be added (90 mL in this example) and
press
Di spl ay:
>XDOS par amet er s
f eed r at e max mL/ mi n
•
Enter the feed rate (in this example 50 mL/min). Accept the value with
Di spl ay:
>XDOS par amet er s
t emp. measur i ng: of f
•
The colon indicates that the
value from a list. Press
Di spl ay:
>XDOS par amet er s
meas. i nt er val 00: 01: 00
•
The time interval at which a measurement is to be added to the list of
measuring points can be determined here. This is used for the
graphics, the list of temperature measurements and the combined
measurement list report. The interval at which the temperature is
measured for controlling the temperature is not dependent on this val-
ue but is carried out every second.
Press
QUIT
NO
J
•
No further entries are necessary in the ‘XDOS parameters’ submenu.
Exit with the
Di spl ay:
par amet er s
>XDOS par amet er s
•
<> is used to open the next submenu.
Di spl ay:
Par amet er
>XDOS dosi ng uni t s
•
Press
Di spl ay:
>XDOS dosi ng uni t s
set up: si ngl e
•
As only 1 Dosino is needed the default value can be confirmed with
Di spl ay:
>XDOS dosi ng uni t s
dr i ve uni t : DOS1
•
The Dosino to be used can be entered with
to the connection number on the 711 Liquino). Accept with
Di spl ay:
>XDOS dosi ng uni t s