Metrohm 711 Liquino User Manual
Page 20

2.5 Connections
711 Liquino, Instructions for use
14
Example of a sample changer method:
>st ar t sequence
1 CTL: Rm: I NI T
>sampl e sequence
1 CTL: Rm: START devi ce1
2 SCN: Rm: r eady1
3 Li f t : 1: shi f t mm
4 CTL: Rm: * * * * * * * * * 1* * * *
5 Wai t 1 s
6 CTL: Rm: * * * * * * * * * 0* * * *
7 Move: 1: sampl e
8 Li f t : 1: wor k mm
9 CTL: Rm: * * * * * * * * * 1* * * *
10 Wai t 1 s
11 CTL: Rm: * * * * * * * * * 0* * * *
12 SCN: Rm: r eady1
13 Move: 1: spec. 1
14 Li f t : 1: wor k mm
15 CTL: Rm: * * * * * * * * * 1* * * *
16 Wai t 1 s
17 CTL: Rm: * * * * * * * * * 0* * * *
18 SCN: Rm: end1
19 St i r : 1: ON s
20 CTL: Rm: START devi ce2
21 SCN: Rm: end2
22 St i r : 1: OFF s
23 Pump: 1. 1: 12 s
24 Pump: 1. * : 10 s
25 Pump: 1. 1: 12 s
28 NOP
>f i nal sequence
1 Li f t : 1: shi f t mm
2 NOP
← Initialize remote interface
← Start Liquino (PIP mode)
← Wait for pipetting step
← Lift out of sample vessel
← Switch pipetting further (corresponds to
— separating bubble is aspirated —
← Reset remote interface
← Position sample beaker
← Immerse pipet tubing in sample
← Switch pipetting further (corresponds to
— sample is aspirated —
← Reset remote interface
← Wait for next pipetting step
← Move to titrating beaker
← Pipet tubing in titrating beaker
← Switch pipetting further (corresponds to
— sample is ejected —
← Reset remote interface
← Wait for end of PIP mode
← Stirrer on
← Start Titrino
← Wait for end of determination
← Switch off stirrer
← Aspirate titration beaker
← Rinse titration beaker
← Aspirate titration beaker
711 Liquino – 731 Relay Box – 772 Pump Unit – Heating Bath
Liquino with 772 Pump Unit as delivery or aspiration pump and a heating
bath.
711
731
772
With the aid of a Metrohm 731 Relay Box the output leads of the 711 Liquino
can be monitored and, according to their line conditions, two low-voltage
d.c. voltage outputs and two mains voltage outputs can be switched.
In this way the heating bath can be switched automatically via the tempera-
ture monitoring function. If the measured temperature falls below a certain
value the bath is heated up.
A connected pump (e.g. the Metrohm 772 Pump Unit) can be selectively
switched on or off during the method procedure.
cable
6.2125.100
Dosinos
heating bath
mains cable