3 Introduction
711 Liquino, Instructions for use
41
•
Accept the entry with
Di spl ay:
>CONT par amet er s
meas. uni t : mol / L
•
Use the
Di spl ay:
>CONT par amet er s
t ar get vol ume 10. 0 mL
•
The target volume corresponds to the volumetric flask volume up to the
mark. In this example enter ‘10’ and confirm this value with
Di spl ay:
>CONT par amet er s
cont ent st d. 1. 000 ppm
•
Enter the standard solution content (in this example 1000 ppm). Press
Di spl ay:
par amet er s
>CONT dosi ng uni t
•
As the ‘>CONT parameters’ submenu contains no further menu items,
the next submenu automatically becomes available. Open it with
Di spl ay:
>CONT dosi ng uni t
dr i ve uni t : DOS1
•
The Dosino to be used can be chosen with
to the connection number on the 711 Liquino). Press
cept the entry.
Di spl ay:
>CONT dosi ng uni t
dos. uni t : def aul t
•
In this menu item the 711 Liquino is informed about the tubing of the
Dosino. Press
Di spl ay:
par amet er s
>CONT r at es
•
All the necessary parameters have now been entered. The parameter
menu is exited with the